PT - JOURNAL ARTICLE AU - Smit, E.J. AU - Vonken, E.-j. AU - Meijer, F.J.A. AU - Dankbaar, J.W. AU - Horsch, A.D. AU - van Ginneken, B. AU - Velthuis, B. AU - van der Schaaf, I. AU - Prokop, M. TI - Timing-Invariant CT Angiography Derived from CT Perfusion Imaging in Acute Stroke: A Diagnostic Performance Study AID - 10.3174/ajnr.A4376 DP - 2015 Oct 01 TA - American Journal of Neuroradiology PG - 1834--1838 VI - 36 IP - 10 4099 - http://www.ajnr.org/content/36/10/1834.short 4100 - http://www.ajnr.org/content/36/10/1834.full SO - Am. J. Neuroradiol.2015 Oct 01; 36 AB - BACKGROUND AND PURPOSE: Timing-invariant (or delay-insensitive) CT angiography derived from CT perfusion data may obviate a separate cranial CTA in acute stroke, thus enhancing patient safety by reducing total examination time, radiation dose, and volume of contrast material. We assessed the diagnostic accuracy of timing-invariant CTA for detecting intracranial artery occlusion in acute ischemic stroke, to examine whether standard CTA can be omitted.MATERIALS AND METHODS: Patients with suspected ischemic stroke were prospectively enrolled and underwent CTA and CTP imaging at admission. Timing-invariant CTA was derived from the CTP data. Five neuroradiologic observers assessed all images for the presence and location of intracranial artery occlusion in a blinded and randomized manner. Sensitivity and specificity of timing-invariant CTA and standard CTA were calculated by using an independent expert panel as the reference standard. Interrater agreement was determined by using κ statistics.RESULTS: We included 108 patients with 47 vessel occlusions. Overall, standard CTA and timing-invariant CTA provided similar high diagnostic accuracy for occlusion detection with a sensitivity of 96% (95% CI, 90%–100%) and a specificity of 100% (99%–100%) for standard CTA and a sensitivity of 98% (95% CI, 94%–100%) and a specificity of 100% (95% CI, 100%–100%) for timing-invariant CTA. For proximal large-vessel occlusions, defined as occlusions of the ICA, basilar artery, and M1, the sensitivity and specificity were 100% (95% CI, 100%–100%) for both techniques. Interrater agreement was good for both techniques (mean κ value, 0.75 and 0.76).CONCLUSIONS: Timing-invariant CTA derived from CTP data provides diagnostic accuracy similar to that of standard CTA for the detection of artery occlusions in acute stroke.