PT - JOURNAL ARTICLE AU - Bhagat, Y.A. AU - Emery, D.J. AU - Naik, S. AU - Yeo, T. AU - Beaulieu, C. TI - Comparison of Generalized Autocalibrating Partially Parallel Acquisitions and Modified Sensitivity Encoding for Diffusion Tensor Imaging DP - 2007 Feb 01 TA - American Journal of Neuroradiology PG - 293--298 VI - 28 IP - 2 4099 - http://www.ajnr.org/content/28/2/293.short 4100 - http://www.ajnr.org/content/28/2/293.full SO - Am. J. Neuroradiol.2007 Feb 01; 28 AB - BACKGROUND AND PURPOSE: Diffusion tensor magnetic resonance imaging (DTI) of the brain is usually acquired with single-shot echo-planar imaging, which is associated with localized signal loss, geometric distortions, and blurring. Parallel imaging can lessen these artifacts by shortening the length of the echo-train acquisition. The self-calibrating parallel acquisition techniques, image domain-based modified sensitivity encoding (mSENSE) and k-space-based generalized autocalibrating partially parallel acquisitions (GRAPPA), were evaluated with DTI of the brain in 5 healthy subjects.METHODS: GRAPPA and mSENSE with higher acceleration factors (R) up to 4 were compared with conventional DTI (with and without phase partial Fourier, another method of reducing the echo-train length) on a 1.5T Sonata scanner (Siemens, Erlangen, Germany). The resulting images and diffusion maps were evaluated qualitatively and quantitatively. Qualitative analysis was performed by 3 reviewers blinded to the technique using image sharpness and the level of artifacts as characteristics for scoring each set of images. Quantitative comparisons encompassed measuring signal-to-noise ratio, Trace/3 apparent diffusion coefficient (ADC), and fractional anisotropy (FA) in 6 white-matter (WM) and gray-matter (GM) regions.RESULTS: Reviewers scored the GRAPPA and mSENSE R = 2 images better than images acquired with conventional techniques. FA contrast was improved at the GM/WM junction in peripheral brain areas. Trace/3 ADC and FA measurements were consistent for all methods. However, R = 3,4 images suffered from reconstruction-related artifacts.CONCLUSIONS: GRAPPA and mSENSE (R = 2) minimized the susceptibility and off-resonance effects associated with conventional DTI methods, yielding high-quality images and reproducible quantitative diffusion measurements.