Length profiles of the microcatheters tested
Excel-14* | Rebar-14 | Prowler-14 | Excelsior | FasTracker-10 | Prowler-10 | |
---|---|---|---|---|---|---|
Proximal marker | 0.59 ± 0.01 | 0.66 ± 0.05 | 1.28 ± 0.04 | 0.60 ± 0.03 | 0.62 ± 0.04 | 1.27 ± 0.03 |
(0.57 ∼ 0.60) | (0.58 ∼ 0.70) | (1.25 ∼ 1.36) | (0.57 ∼ 0.63) | (0.58 ∼ 0.65) | (1.25 ∼ 1.36) | |
Distal marker | 0.61 ± 0.01 | 0.93 ± 0.02 | 1.41 ± 0.05 | 0.59 ± 0.03 | 0.61 ± 0.02 | 1.44 ± 0.07 |
(0.59 ∼ 0.62) | (0.90 ∼ 0.95) | (1.33 ∼ 1.47) | (0.56 ∼ 0.63) | (0.60 ∼ 0.62) | (1.34 ∼ 1.52) | |
Radiographically transparent zone of the catheter tip | 0.80 ± 0.05 | 0.82 ± 0.10 | 0.50 ± 0.09 | 0.74 ± 0.04 | 0.79 ± 0.12 | 0.44 ± 0.04 |
(0.75 ∼ 0.88) | (0.73 ∼ 0.93) | (0.43 ∼ 0.65) | (0.68 ∼ 0.77) | (0.65 ∼ 0.88) | (0.40 ∼ 0.50) | |
Position of the coil detachment | 1.50 ± 0.23 | 1.18 ± 0.21 | 1.43 ± 0.15 | 1.56 ± 0.27 | 1.33 ± 0.15 | 1.42 ± 0.17 |
zone from the catheter end with usual alignment | (1.22 ∼ 1.73) | (0.08 ∼ 1.29) | (1.24 ∼ 1.57) | (1.22 ∼ 1.80) | (1.21 ∼ 1.50) | (1.18 ∼ 1.57) |
Note.—Average (mm) ± SD (minimum ∼ maximum); usual alignment refers to the marker alignment with which non-SyngerG GDC is advanced until the proximal end of the radiopaque marker on the delivery wire is exactly distal to the proximal marker on a microcatheter.
* An Excel-14 microcatheter with a long transparent zone of the catheter tip (1.68 mm) is excluded.