Table 1:

Conditions used to expose the programmable valves to the 3T MR system

TestPressure SettingOrientationArrow Direction
10SagittalUp
20SagittalDown
30AxialUp
40AxialDown
50CoronalIn
60CoronalOut
78SagittalUp
88SagittalDown
98AxialUp
108AxialDown
118CoronalIn
128CoronalOut
1320SagittalUp
1420SagittalDown
1520AxialUp
1620AxialDown
1720CoronalIn
1820CoronalOut
195Simulated patientIn
205Simulated patientOut
  • Note:—All conditions were in-out × 5. There were no changes under any conditions.