Table 2:

Subjective evaluation for the 2D DSA regarding the depiction of simulated aneurysms

Contrast Material Concentration & Dose ModeScore for Image QualityP
54321Mean ± SD*
300 mgI/mL
    Standard-dose mode of FPD system5700005.00 ± 0.00**.038
    Low-dose mode of FPD system5700005.00 ± 0.00**.038
    Conventional-dose mode of I.I.-TV system5232004.88 ± 0.43
150 mgI/mL
    Standard-dose mode of FPD system5511004.95 ± 0.29**<.001
    Low-dose mode of FPD system5241004.90 ± 0.36**<.001
    Conventional-dose mode of I.I.-TV system41105104.60 ± 0.73
100 mgI/mL
    Standard-dose mode of FPD system41124004.65 ± 0.61**<.001
    Low-dose mode of FPD system39135004.60 ± 0.65**<.001
    Conventional-dose mode of I.I.-TV system115261323.00 ± 0.85
  • Note:—DSA indicates digital subtraction angiography; FPD, flat panel detector; I.I.-TV, image intensifier television. Data are mean ± SD of scores for image quality evaluated using a 5-point scale (1, clearly visualized; 2, good; 3, reasonably good; 4, poor; 5, very poor)

  • ** Significantly higher to I.I.-TV system (the Wilcoxon signed rank test).