Diagnostic performance according of each observer regarding the detection of the simulated aneurysmal blebs
Contrast Material Concentration & Dose Mode | Observer | Mean Az | ||||||
---|---|---|---|---|---|---|---|---|
1 | 2 | 3 | 4 | 5 | 6 | 7 | ||
300 mgI/mL | ||||||||
Standard-dose mode of FPD system | 0.93 | 0.91 | 0.89 | 0.92 | 0.91 | 0.92 | 0.94 | 0.92 |
Low-dose mode of FPD system | 0.89 | 0.89 | 0.91 | 0.92 | 0.92 | 0.94 | 0.95 | 0.92 |
Conventional-dose mode of I.I.-TV system | 0.92 | 0.91 | 0.86 | 0.98 | 0.87 | 0.95 | 0.90 | 0.91 |
150 mgI/mL | ||||||||
Standard-dose mode of FPD system | 0.81 | 0.79 | 0.83 | 0.83 | 0.83 | 0.80 | 0.82 | 0.82 |
Low-dose mode of FPD system | 0.83 | 0.79 | 0.85 | 0.82 | 0.83 | 0.85 | 0.84 | 0.83 |
Conventional-dose mode of I.I.-TV system | 0.80 | 0.78 | 0.85 | 0.81 | 0.82 | 0.84 | 0.82 | 0.82 |
100 mgI/mL | ||||||||
Standard-dose mode of FPD system | 0.84 | 0.88 | 0.87 | 0.80 | 0.93 | 0.87 | 0.80 | 0.86* |
Low-dose mode of FPD system | 0.84 | 0.88 | 0.86 | 0.79 | 0.93 | 0.87 | 0.79 | 0.85* |
Conventional-dose mode of I.I.-TV system | 0.67 | 0.55 | 0.54 | 0.57 | 0.55 | 0.54 | 0.52 | 0.56 |
Note:—Az indicates areas under the receiver operating characteristic curve; FPD, flat panel detector; I.I.-TV, image intensifier television.
* Significantly higher to I.I.-TV system (P < .01).