Computational model-based calculations of intra-aneurysmal hemodynamics before and after flow-diversion treatment with the use of patient-specific boundary conditions from the dual-sensor guidewire
Measure | Pretreatment | Posttreatment | % Change | P | |
---|---|---|---|---|---|
Blood flow, mL/min | Time-averaged | 81.36 | 51.38 | −39.33% | .07 |
Peak systolic | 123.25 | 88.28 | −28.84% | .07 | |
WSS, Pa | Time-averaged | 1.99 | 0.92 | −56.77% | .03 |
Peak systolic | 3.92 | 2.20 | −51.89% | .01 | |
WSSG, Pa/m | Time-averaged | 2807.95 | 1730.70 | −43.38% | .06 |
Peak systolic | 6261.35 | 4236.05 | −42.93% | .04 |