Table 1:

Sequence parameters

ParameterCartesian T2WI TSEMultiVane T2WI TSESpiral T2WI TSE
Technique2D TSE2D TSE2D TSE
FOV AP/FH/RL (mm)230 × 230 × 149 230 × 230 × 149 230 × 230 × 149 
Acquired voxel size (mm3)0.7 × 0.7 × 4.0 0.7 × 0.7 × 4.0 0.7 × 0.7 × 4.0 
Reconstructed voxel size (mm3)0.45 × 0.45 × 4.0 0.45 × 0.45 × 4.0 0.45 × 0.45 × 4.0
No. of slices303030
TR (ms)5300 5300 5300
TE equivalent (ms)90 90 90 
Flip angle90°90°90°
Refocusing flip angle180°180°180°
TSE factor28308
Spiral acq windowSpiral in-out, 11.4 ms
Parallel imagingNoSENSE 1.5No
NSA212
Receiver bandwidth (Hz/pixel)560 290 114 
SAR (W/kg)<1.9<2.0<1.6
dB/dT (T/s)523372
Acquisition time (min:sec)04:1402:4902:12
  • Note:—NSA indicates number of signal averages; acq, acquisition; SAR, specific absorption rate; dB/dT, ratio between the amount of change in amplitude of the magnetic field (dB) and the time it takes to make that change (dt); T/s, Tesla/second; AP/FH/RL, anterior-posterior/foot-head/right-left; –, scan parameter does not exist for this sequence.