Table 1:

Scan details

Canon Aquilion ONEPhilips IQonPrototype DL-CBCT
Tube (kV)100120120
Tube current (mAs/mA) (average)196 (Auto-modulation)114 (Auto-modulation)310
Scan time/rotation time (sec)3.3/0.5 (Full rotation)2.5/0.3 (Full rotation)20.0/20.0 (200° rotation)
Nominal beam width (mm)80 × 0.50064 × 0.625194.700
Pitch factor0.8130.671NA
Display FOV coronal × sagittal × axial (mm3)210.9 × 210.9 × 160.0210.0 × 210.0 × 160.0251.8 × 251.8 × 194.7
Section thickness (mm)0.500.670.66
Matrix size512 × 512512 × 512384 × 384
Reconstruction kernelFC43Filter UAStåhl et al15a
Reconstruction algorithmAIDR 3D eStandardiDose4 level 4Ståhl et al15a
Avg CTDIvol (16-cm phantom)20.0 mGy21.3 mGyNA
Air kerma (in an 18-cm water phantom)bNANA57.6 mGy
MTFc50%: 3.7850%: 3.4650%: 3.57
10%: 6.5710%: 6.6510%: 6.04
  • Note:—CTDIvol indicates volume CT dose index; MTF, modulation transfer function; NA, not applicable.

  • a Details of the prototype algorithm are described in Ståhl et al.15

  • b Air kerma in an 18-cm diameter plastic water phantom at the center of the scan length, measured in accordance with American Association of Physicists in Medicine Task Group Report 111.26

  • c Generated from consecutive scans of the upper bead of a Catphan CTP528 module (The Phantom Laboratory).