Typical optimized whole-brain DTI acquisition parameters in a 1.5T or 3T MR imaging system*

Acquisition Parameters3T1.5T
Spatial resolution2.0 × 2.0 × 2.0 mm2.5 × 2.5 × 2.5 mm
Acquisition matrix128 × 128 × 6096 × 96 × 50
FOV256 mm240 mm
No. DWIs3030
No. minimally weighted images55
No. repetitions (NEX)11
b-value1000 s/mm21000 s/mm2
TE/TR70 ms/<12 seconds70 ms/<10 seconds
Total acquisition time<8 minutes<7 minutes
  • * The hardware is assumed to be equipped with an 8-channel head coil and gradients of 40 mT/m. It is also assumed that parallel acquisition is done with a SENSE reduction factor of 2, partial k-space acquisition of 62.5% (only 5/8 of the phase-encoding lines after a reduction to half by parallel acquisition is acquired), and an interleaved multisection SS-EPI sequence with no gap.