Table 2:

Diagnostic performance color duplex ultrasonography, B-flow imaging, and a combination of these tests in detecting 70% of 99% internal carotid artery stensosis

CriteriaSens (%)Spec (%)PPV (%)NPV (%)ACC (%)MRate (%)
BFI ≥ 70%65989192928.1
PSV ≥ 230 cm/s91937498927.6
PSV ≥ 230 cm/s + BFI ≥ 70%90989098972.8
EDV ≥ 75 cm/s91958198945.8
EDV ≥ 75 cm/s + BFI ≥ 70%95999599991.4
PSVICA/CCA ≥ 3.694968399954.7
PSVICA/CCA ≥ 3.6 + BFI ≥ 70%95999599991.4
  • Note.—Sens indicates sensitivity; Spec, specificity; PPV, positive predictive value; NPV, negative predictive value; Acc, accuracy; MRate, misclassification rate; BFI, B-flow imaging; PSV, peak systolic velocity; EDV, end diastolic velocity; PSVICA/CCA, ratio of ICA to common carotid artery peak systolic velocity.