Table 2:

Statistical results comparing the CBF values from CT and CBCT from different scan sets for the 6-mL/s 50% contrast-concentration injection protocola

Scan SetCCC (ρ)Correlation Coefficient (r)Difference (CT-CBCT) (mL/100 g/min)Bland-Altman 95% Limits
MeanSD
B-D-F0.7960.876.568.34−9.78–22.89
A-B-C-D-E-F0.7720.888.607.65−6.40–23.59
A-C0.7510.888.787.68−6.27–23.84
  • a The top 3 candidates are shown.