Summary of the essential imaging parameters for PICORE and FAIR ASL sequences

PICORE Q2TIPSFAIR-QUIPSS II
Matrix64 × 6464 × 64
No. of sections2024
Pixel size (mm2)3.44 × 3.443.44 × 3.44
Section thickness (mm)64
Gap (mm)10
FOV220220
Parallel imaging (GRAPPA)22
Phase, partial Fourier7/86/8
Section orderAscendingAscending
Section orientationACPC alignedACPC aligned
Fat saturationYesYes
Coil-combination algoSum of squaresSum of squares
Bandwidth2604 Hz/Px2368 Hz/Px
Echo-spacing0.470.49
Saturation slabParallel feet, gap = 25.6 mm, thickness = 110 mmParallel feet/head gap = 10 mm, thickness = 100 mm
TE (ms)2112
TR (ms)50004000
TI 1 (ms)700600
TI 2 (ms)18001600
Saturation stop time (ms)1600
Flow limit (cm/s)5100
No. of images71105
Total acquisition time6 Minutes 12 seconds7 Minutes 6 seconds
Mo reference imageNoYes
  • Note:—ACPC indicates anterior/posterior commissure; algo, algorithm; Q2TIPS, quantitative imaging of perfusion with a single subtraction with thin-section TI1 periodic saturation; QUIPSS II, quantitative imaging of perfusion using a single subtraction II; Mo, steady-state magnetization; GRAPPA, generalized autocalibrating partially parallel acquisitions.