Table 1:

Scan protocols and parameters used including CTDIvol, DLP, and effective dose

Dose ProtocolReconstruction TechniqueKernelkVmAsPitchRotation Time (s)CTDIvol (mGy)DLP (mG × cm)Effective Dose (mSv)
Emergency protocol headASIR-50Bone120314–48810.647.551141.264.9
Reference dose protocolFBPBone, std1201000.5136.69930.783.5
ASIR-50Bone, std
ASIR-100Bone, std
MBIRstd
Low-dose protocol IFBPBone, std100350.50.54.18106.180.4
ASIR-50Bone, std
ASIR-100Bone, std
MBIRstd
Low-dose protocol IIFBPBone, std80400.50.52.6466.690.3
ASIR-50Bone, std
ASIR-100Bone, std
MBIRstd
Low-dose protocol IIIFBPBone, std80150.50.50.9925.130.1
ASIR-50Bone, std
ASIR-100Bone, std
MBIRstd
Low-dose protocol IVFBPBone, std80100.50.40.5313.390.1
ASIR-50Bone, std
ASIR-100Bone, std
MBIRstd
  • Note:—DLP indicates dose-length product; std, standard.