Number of discrepancies between sCTA and DSA for both readers

Intracranial Aneurysm FeatureDiscrepancies between sCTA and DSAComments
Reader 1Reader 2
Aneurysm closure (Raymond 1)00NA
Complete device deployment01Device deployment was scored complete on sCTA but incomplete on DSA; a second reading of this case confirmed complete device deployment on DSA
Complete wall apposition11In 1 patient, both readers scored wall apposition incomplete on sCTA but complete on DSA; a second reading of this case confirmed incomplete wall apposition on DSA
Complete neck coverage00NA
Vessel patency distal (nonstenotic)10Vessel patency was scored nonstenotic on sCTA but stenotic on DSA in 1 case; a second reading confirmed nonstenosis on DSA
Vessel patency proximal (nonstenotic)00NA
Side branch patency (no change)01In 1 case, a side branch was missed on follow-up sCTA; a second reading of the case confirmed a patent side branch on sCTA
  • Note:—NA indicates not applicable.