Epilepsy-dedicated MR imaging protocola

MR Imaging SequenceNo. of Slices/Thickness (mm)Voxel Size (mm3)TI/TR/TE/α (ms/ms/ms/degree)Acquisition Time (Min:Sec)
Sag 3D-MPRAGE160/11 × 1 × 1900/2000/2.26/124:40
Sag 3D-FLAIR-SPACE160/11 × 1 × 11800/5000/388/var6:52
Ax 2D-T2-TSE42/30.4 × 0.4 × 35040/102/1504:34
Ax 2D-T2*23/50.7 × 0.7 × 5639/19.9/202:33
Cor 2D-T2-STIR40/20.4 × 0.4 × 2100/5390/25/1408:07
Cor 2D FLAIR68/20.7 × 0.7 × 0.22500/9000/87/1504:14
Ax 2D-DWI-SE EPI23/50.6 × 0.6 × 53400/850:46
Sag 3D-MP2RAGE192/11 × 1 × 1700, 2200/2000/2.26/128:52
  • Note:SPACE indicates sampling perfection with application-optimized contrasts by using flip angle evolution; SE, spin-echo; α, flip angle; Sag, sagittal; Ax, axial; Cor, coronal; var, variable flip angle.

  • a 3T Magnetom Prisma (Siemens).