Table 2:

Diagnostic performance of CT perfusion and DCE-MR imaging in the prediction of wound failure after head and neck free flap reconstruction

Blood Flow (mL/100 mL/min)Blood Volume (mL/100 mL)MTT (Sec)tMIPVpKtrans (min−1)Peak EnhancementTME (Sec)
Cutoff17.35.6216.110.70.060.15155113
Sensitivity111110.8011
Specificity0.940.880.880.940.810.940.810.69
PPV0.750.600.600.750.630.800.630.50
NPV111110.9411
Accuracy0.950.900.900.950.860.910.860.76
AUC0.960.960.960.980.920.820.910.81
  • Note:—PPV indicates positive predictive value; NPV, negative predictive value; AUC, area under the curve.