Skip to main content
Advertisement

Main menu

  • Home
  • Content
    • Current Issue
    • Accepted Manuscripts
    • Article Preview
    • Past Issue Archive
    • AJNR Case Collection
    • Case of the Week Archive
    • Classic Case Archive
    • Case of the Month Archive
  • Special Collections
    • Spinal CSF Leak Articles (Jan 2020-June 2024)
    • 2024 AJNR Journal Awards
    • Most Impactful AJNR Articles
  • Multimedia
    • AJNR Podcast
    • AJNR Scantastics
    • Video Articles
  • For Authors
    • Submit a Manuscript
    • Author Policies
    • Fast publishing of Accepted Manuscripts
    • Graphical Abstract Preparation
    • Manuscript Submission Guidelines
    • Imaging Protocol Submission
    • Submit a Case for the Case Collection
  • About Us
    • About AJNR
    • Editorial Board
  • More
    • Become a Reviewer/Academy of Reviewers
    • Subscribers
    • Permissions
    • Alerts
    • Feedback
    • Advertisers
    • ASNR Home
  • Other Publications
    • ajnr

User menu

  • Alerts
  • Log in

Search

  • Advanced search
American Journal of Neuroradiology
American Journal of Neuroradiology

American Journal of Neuroradiology

ASHNR American Society of Functional Neuroradiology ASHNR American Society of Pediatric Neuroradiology ASSR
  • Alerts
  • Log in

Advanced Search

  • Home
  • Content
    • Current Issue
    • Accepted Manuscripts
    • Article Preview
    • Past Issue Archive
    • AJNR Case Collection
    • Case of the Week Archive
    • Classic Case Archive
    • Case of the Month Archive
  • Special Collections
    • Spinal CSF Leak Articles (Jan 2020-June 2024)
    • 2024 AJNR Journal Awards
    • Most Impactful AJNR Articles
  • Multimedia
    • AJNR Podcast
    • AJNR Scantastics
    • Video Articles
  • For Authors
    • Submit a Manuscript
    • Author Policies
    • Fast publishing of Accepted Manuscripts
    • Graphical Abstract Preparation
    • Manuscript Submission Guidelines
    • Imaging Protocol Submission
    • Submit a Case for the Case Collection
  • About Us
    • About AJNR
    • Editorial Board
  • More
    • Become a Reviewer/Academy of Reviewers
    • Subscribers
    • Permissions
    • Alerts
    • Feedback
    • Advertisers
    • ASNR Home
  • Follow AJNR on Twitter
  • Visit AJNR on Facebook
  • Follow AJNR on Instagram
  • Join AJNR on LinkedIn
  • RSS Feeds

Welcome to the new AJNR, Updated Hall of Fame, and more. Read the full announcements.


AJNR is seeking candidates for the position of Associate Section Editor, AJNR Case Collection. Read the full announcement.

 

Research ArticleAdult Brain
Open Access

Clinical Experience of 1-Minute Brain MRI Using a Multicontrast EPI Sequence in a Different Scan Environment

K.H. Ryu, H.J. Baek, S. Skare, J.I. Moon, B.H. Choi, S.E. Park, J.Y. Ha, T.B. Kim, M.J. Hwang and T. Sprenger
American Journal of Neuroradiology March 2020, 41 (3) 424-429; DOI: https://doi.org/10.3174/ajnr.A6427
K.H. Ryu
aFrom the Department of Radiology (K.H.R., H.J.B., J.I.M., B.H.C., S.E.P., J.Y.H., T.B.K.), Gyeongsang National University School of Medicine and Gyeongsang National University Changwon Hospital, Changwon, Republic of Korea
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for K.H. Ryu
H.J. Baek
aFrom the Department of Radiology (K.H.R., H.J.B., J.I.M., B.H.C., S.E.P., J.Y.H., T.B.K.), Gyeongsang National University School of Medicine and Gyeongsang National University Changwon Hospital, Changwon, Republic of Korea
bDepartment of Radiology (H.J.B.), Institute of Health Sciences, Gyeongsang National University School of Medicine, Jinju, Republic of Korea
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for H.J. Baek
S. Skare
cDepartment of Clinical Neuroscience (S.S., T.S.), Karolinska Institute, Stockholm, Sweden
dDepartment of Neuroradiology (S.S.), Karolinska University Hospital, Stockholm, Sweden
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for S. Skare
J.I. Moon
aFrom the Department of Radiology (K.H.R., H.J.B., J.I.M., B.H.C., S.E.P., J.Y.H., T.B.K.), Gyeongsang National University School of Medicine and Gyeongsang National University Changwon Hospital, Changwon, Republic of Korea
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for J.I. Moon
B.H. Choi
aFrom the Department of Radiology (K.H.R., H.J.B., J.I.M., B.H.C., S.E.P., J.Y.H., T.B.K.), Gyeongsang National University School of Medicine and Gyeongsang National University Changwon Hospital, Changwon, Republic of Korea
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for B.H. Choi
S.E. Park
aFrom the Department of Radiology (K.H.R., H.J.B., J.I.M., B.H.C., S.E.P., J.Y.H., T.B.K.), Gyeongsang National University School of Medicine and Gyeongsang National University Changwon Hospital, Changwon, Republic of Korea
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for S.E. Park
J.Y. Ha
aFrom the Department of Radiology (K.H.R., H.J.B., J.I.M., B.H.C., S.E.P., J.Y.H., T.B.K.), Gyeongsang National University School of Medicine and Gyeongsang National University Changwon Hospital, Changwon, Republic of Korea
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for J.Y. Ha
T.B. Kim
aFrom the Department of Radiology (K.H.R., H.J.B., J.I.M., B.H.C., S.E.P., J.Y.H., T.B.K.), Gyeongsang National University School of Medicine and Gyeongsang National University Changwon Hospital, Changwon, Republic of Korea
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for T.B. Kim
M.J. Hwang
eMR Applications and Workflow, GE Healthcare (M.J.H.), Seoul, Republic of Korea
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for M.J. Hwang
T. Sprenger
cDepartment of Clinical Neuroscience (S.S., T.S.), Karolinska Institute, Stockholm, Sweden
fMR Applied Science Laboratory Europe (T.S.), GE Healthcare Stockholm, Sweden.
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for T. Sprenger
  • Article
  • Figures & Data
  • Supplemental
  • Info & Metrics
  • Responses
  • References
  • PDF
Loading

Abstract

BACKGROUND AND PURPOSE: The long scan time of MR imaging is a major drawback limiting its clinical use in neuroimaging; therefore, we aimed to investigate the clinical feasibility of a 1-minute full-brain MR imaging using a multicontrast EPI sequence on a different MR imaging scanner than the ones previously reported.

MATERIALS AND METHODS: We retrospectively reviewed the records of 146 patients who underwent a multicontrast EPI sequence, including T1-FLAIR, T2-FLAIR, T2WI, DWI, and T2*WI sequences. Two attending neuroradiologists assessed the image quality of each sequence to compare the multicontrast EPI sequence with routine MR imaging protocols. We used the Wilcoxon signed rank test and McNemar test to compare the 2 MR imaging protocols.

RESULTS: The multicontrast EPI sequence generally showed sufficient image quality of >2 points using a 4-point assessment scale. Regarding image quality and susceptibility artifacts, there was no significant difference between the multicontrast EPI sequence DWI and routine DWI (P > .05), attesting to noninferiority of the multicontrast EPI, whereas there were significant differences in the other 4 sequences between the 2 MR imaging protocols.

CONCLUSIONS: The multicontrast EPI sequence showed sufficient image quality for clinical use with a shorter scan time; however, it was limited by inferior image quality and frequent susceptibility artifacts compared with routine brain MR imaging. Therefore, the multicontrast EPI sequence cannot completely replace the routine MR imaging protocol at present; however, it may be a feasible option in specific clinical situations such as screening, time-critical diseases or for use with patients prone to motion.

ABBREVIATIONS:

EPIMix
multicontrast EPI sequence
SWAN
susceptibility-weighted angiography

The long scan time of brain MR imaging is a major drawback limiting its clinical use. To shorten the scan time, fast imaging techniques have been introduced.1⇓⇓⇓⇓-6 However, most studies have examined specific clinical conditions, thus only evaluating limited sequences.7⇓⇓⇓⇓⇓⇓⇓-15 Another fast technique, synthetic MR imaging, quantitatively approaches the absolute physical properties for single-scan, multiple-contrast generation with a total scan time of 4–5 minutes.4⇓-6,16,17 Routine clinical use of synthetic MR imaging is limited because it cannot produce DWI and T2*WI and provide inferior image quality in synthetic T2-FLAIR due to partial volume effects.18 Following technical advances, some studies have focused on ultrafast MR imaging with 5 essential sequences in patients prone to motion, combining advanced imaging techniques with optimized accelerated versions of commercially available sequences, and the studies showed sufficient image quality for diagnostic use with scan times of 2 minutes 47 seconds and 4 minutes.19,20

Recently, a new 1-minute full brain, multicontrast EPI sequence (EPIMix) was introduced,21 which can shorten scan time and generate T1-FLAIR, T2WI, T2-FLAIR, DWI, apparent diffusion coefficient, and T2*WI sequences in 78 seconds. Only 1 study showed comparable diagnostic performance between EPIMix and routine brain MR imaging for clinical use.22 However, the EPIMix studies did not fully evaluate image quality according to individual sequences, and they acquired images using the same MR imaging scanners at the same institution.21,22 Therefore, here, we evaluated the clinical feasibility of EPIMix on a different MR imaging scanner by comparing EPIMix with the routine MR imaging protocol. We hypothesized that EPIMix would be noninferior to the routine protocol for clinical use with a 10-fold reduction in scan time to obtain the basic 5 sequences.

MATERIALS AND METHODS

Patients

We retrospectively reviewed the database of our institution and identified 146 patients who underwent diagnostic brain MR imaging examinations, including EPIMix and routine protocols with the 5 basic sequences in a single examination session from January 2019 to May 2019. The included patients were 75 men and 71 women (age range, 15–89 years; mean age, 57.2 years). MR imaging examinations were performed for headache (32/146, 21.9%), dizziness or vertigo (29/146, 19.9%), brain metastasis work-up (17/146, 11.6%), follow-up of treated intracranial aneurysm (10/146, 6.8%), weakness of the extremities (7/146, 4.8%), syncope (7/146, 4.8%), brain tumor follow-up (7/146, 4.8%), sensory change (6/146, 4.1%), altered mental status (6/146, 4.1%), parkinsonism (5/146, 3.4%), infarction follow-up (5/146, 3.4%), intracranial hemorrhage (3/146, 2.1%), dysarthria (2/146, 1.4%), and other reasons (10/146, 6.8%).

This study was approved by the institutional review board at Gyeongsang National University Changwon Hospital. The institutional review board determined that patient approval and informed consent were not required due to the retrospective nature of the study.

Imaging Acquisition

MR imaging was performed using a 3T system (Signa Architect; GE Healthcare, Milwaukee, Wisconsin) with a 48-channel head coil. Both the EPIMix and routine MR imaging protocols included 5 basic image contrasts: T1-FLAIR, T2WI, T2-FLAIR, DWI, and susceptibility-weighted angiography (SWAN)/T2*WI. At our institution, we use the ultrafast protocol for clinical purposes in specific clinical situations (such as motion-prone patients or patients with time-critical disease), and we routinely acquire images using the 2 sets of MR imaging protocols (EPIMix and routine protocols) in such clinical cases.

EPIMix Brain MR Imaging Protocol

The EPIMix pulse sequence was acquired using a Discovery 750w 3T or an Optima 450 1.5T MR imaging system (GE Healthcare) in the original study.21 Compared with the initial version of EPIMix,21 the present version was improved by reducing the wait time before the T2-FLAIR block, and 2 EPI echoes were applied for DWI and T2-FLAIR to increase the SNR, with the same version as that used in a recent study.22 The present version of EPIMix was optimized for our MR imaging system; therefore, EPIMix was acquired using a section thickness of 5 mm, in-plane acceleration factor of R = 3, and section number of 28. The EPIMix protocol is detailed in the Table. It included axial T1-FLAIR, T2WI, T2-FLAIR, DWI, and T2*WI. More detailed technical aspects of EPIMix can be found in a previous study.21 The net scan time was 1 minute 12 seconds, and the total scan time was 1 minute 28 seconds.

View this table:
  • View inline
  • View popup

Acquisition parameters of EPIMix and routine protocols

Routine Brain MR Imaging Protocol

The routine brain MR imaging protocol included 5 basic sequences of axial T1-FLAIR, T2WI, T2-FLAIR, DWI, and SWAN. Details of routine MR imaging protocols can be found in the Table. The net scan time of routine MR imaging was 10 minutes 44 seconds, and the total scan time was 11 minutes 59 seconds.

Image Analysis

All datasets were anonymized with randomization, and 2 readers, blinded to patient information, reviewed all images using a PACS. Two attending neuroradiologists with 9 and 4 years of experience, respectively, performed an independent analysis of both brain MR imaging protocols to evaluate the image quality of EPIMix from a clinical feasibility perspective. On-line Figure 1 shows a representative example of the images obtained with the 2 different MR imaging protocols. For interpretation, each reader first analyzed and assessed all EPIMix sequences. After a 4-week memory-washout period, the readers analyzed and assessed all routine MR imaging sequences. The evaluation included assessments of overall image quality, visualization of several anatomic structures, and the severity of artifacts.

Fig 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig 1.

A 59-year-old woman with intracranial hemorrhage. A right thalamic hemorrhage with intraventricular extension to the right lateral ventricle shows mainly hyperintensity on T1-FLAIR (A and G), hyperintensity on T2WI (B and H) and T2-FLAIR (C and I), blooming artifacts on T2*WI (D) and SWAN (J), and diffusion restriction on DWI (E and K) (arrows in each sequence). These findings are well-visualized on both EPIMix MR imaging (A–F) and routine MR imaging (G–L).

For T1-FLAIR, T2WI, and T2-FLAIR, the image quality was rated according to the following 4 criteria: 1, overall image quality; 2, differentiation of gray matter–white matter at the level of the lateral ventricles; 3, demarcation of the basal ganglia at the level of the foramen of Monro; and 4, demarcation of the sulci. For DWI and T2*WI/SWAN, the overall image quality was rated. Each criterion was graded on a Likert scale of 1–4 points:23 1, inadequate (not acceptable for diagnostic use); 2, sufficient (acceptable for diagnostic use but with minor issues); 3, good (acceptable for diagnostic use); and 4, excellent (acceptable for diagnostic use).

The severity of artifacts was rated for all sequences according to the following 2 criteria: 1, motion artifacts; 2, susceptibility artifacts. Each criterion was also scored on a 4-point Likert scale, in which23 images contain the following: 1, severe artifacts (not acceptable for diagnostic use; 2, moderate artifacts (sufficient for diagnostic use but with minor issues); 3, mild artifacts (acceptable for diagnostic use because minor artifacts do not adversely affect diagnostic use); and 4, images do not contain visible artifacts (acceptable for diagnostic use). Then, the readers’ ratings were simply dichotomized as follows: 1) image quality assessments, sufficient for clinical use (2 points) versus excellent for clinical use (≥3 points); and 2) artifact assessments, severe degree (2 points) versus lesser degree (≥3 points) because there was no case assigned 1 point in the image quality assessment.

Statistical Analysis

We assigned numeric values to the image quality assessments. Although we did not directly statistically compare the mean values of the readers’ ratings because they did not strictly comprise continuous variables, we decided to present a summary of the readers’ ratings for each MR imaging sequence, expressed as mean ± SD. We used the Wilcoxon signed rank test to compare ordinal variables and the McNemar test to compare dichotomous variables between the 2 MR imaging protocols. We also calculated the disagreement percentage of the image quality between the proposed EPIMix protocol and the routine protocol. Additionally, we present the interobserver agreement as the percentage agreement. We performed all statistical analyses with statistical software (SPSS, Version 24.0; IBM, Armonk, New York), and we considered P values < .05 statistically significant.

Data Availability Statement

The anonymized data of this study will be shared on request by any qualified investigator.

RESULTS

Study Population

Of 146 patients, 117 (80.1%) had abnormal MR imaging findings and 29 (19.9%) had findings that were considered normal. There were the following diagnoses: ischemic/hemorrhagic stroke or small-vessel disease (79/146, 54.1%), intracranial neoplasm (9/146, 6.2%), vascular abnormality (9/146, 6.2%), infectious or demyelinating disease (5/146, 3.4%), metabolic or degenerative disorder (4/146, 2.7%), congenital CNS malformation (1/146, 0.7%), and miscellaneous, including trauma or indeterminate condition (10/146, 6.8%).

Image Quality Assessment

The On-line Table provides the mean scores from the 2 readers, the interobserver percentage agreement of the readers’ ratings, and disagreement percentages of the dichotomized mean scores for the 5 sequences of the 2 MR imaging protocols. Although the overall image quality of all sequences derived from EPIMix, except for DWI, showed a significant difference (P < .05) between the EPIMix and the routine protocols, the EPIMix sequence showed at least sufficient image quality, with an assessment rating  of >2 points on average (Figs 1–3 and On-line Figs 1–2). Among the 5 EPIMix sequences, T2WI had the lowest mean score in the overall image quality assessment. In contrast, the overall quality of EPIMix DWI was not significantly different from that of routine DWI (P > .05). EPIMix T1-FLAIR, T2WI, T2-FLAIR, and T2*WI showed significant susceptibility artifacts compared with the corresponding sequences of routine MR imaging (P < .05). However, fewer motion artifacts tended to be observed in the EPIMix than in the routine protocol. Additionally, there was no difficulty in detecting small lesions on EPIMix in the enrolled patients (Fig 3).

Fig 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig 2.

A 48-year-old woman with brain abscesses. There are 2 irregularly shaped masslike lesions with perilesional edema in the right temporal lobe. The internal content shows hypointensity on T1-FLAIR (A and G) and hyperintensity on T2WI (B and H) and T2-FLAIR (C and I) (arrows). The internal content shows diffusion restriction on DWI (E and K) (arrows). There are blooming artifacts on T2*WI (D) and SWAN (J) (arrows). These findings are well-visualized on both EPIMix MR imaging (A–F) and routine MR imaging (G–L).

Fig 3.
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig 3.

A 52-year-old man with Wernicke encephalopathy. Signal changes in the tectal plate of the midbrain and periaqueductal gray matter showing hypointensity on T1-FLAIR (A and G), hyperintensity on T2WI (B and H) and T2-FLAIR (C and I), and hyperintensity on DWI (E and K) are well-visualized on both EPIMix MR imaging (A–F) and routine MR imaging (G–L) (arrows in each sequence).

DISCUSSION

In the current study, we assessed the diagnostic image quality of the EPIMix sequence outside the institution that developed it using a different scan environment and found that the EPIMix protocol had sufficient image quality with fewer motion artifacts with a short scan time of approximately 1 minute. However, the EPIMix protocol was inferior in terms of overall image quality, with more severe susceptibility artifacts than those observed in routine brain MR imaging.

Previous clinical fast brain MR imaging protocol studies have shown comparable image quality and high diagnostic concordance with conventional MR imaging.19,20 However, the scan time in those studies varied, ranging from 2 to 10 minutes, being much longer than that of the present EPIMix protocol.14,15,19 To date, the fastest brain MR imaging protocol reported required 1 minute 7 seconds of net scan time; however, the total scan time was 2 minutes 47 seconds due to the prescanning time before each sequence,20 whereas the EPIMix protocol, including all 5 essential sequences, required only a single prescan at the start of the MR imaging examination. Moreover, the EPI readout is more SNR-efficient than the corresponding fast spin-echo readout, and the smaller matrix size can accelerate the EPIMix scan.21 On the basis of these advantages, the total EPIMix scan time appears to be the shortest among those of the reported ultrafast protocols.

The EPIMix protocol uses different magnetization preparations combined with 1 of 2 sets of EPI readout trains to directly acquire the weighted contrast in a short scan time of 1 minute.21 It also uses a dynamic combination of 11 sequence modules with a unique role in the acquisition to generate a specific image contrast, and 2 EPI readout modules are reused throughout the sequence and are prepended by other modules to form the desired MR contrasts.21 This new 1-minute EPIMix protocol showed diagnostic performance comparable with that of routine brain MR imaging in terms of disease identification and categorization in a previous clinical study;22 however, the authors did not analyze image quality and characteristics on the basis of individual sequences derived from EPIMix.22 From the perspective of feasibility, analyses of the image quality and characteristics of each image sequence are essential to apply EPIMix to clinical use to cover a wide range of patient groups because important image sequences for diagnosis differ depending on pathology. Therefore, we performed image quality analysis for individual image sequences to evaluate the clinical feasibility of EPIMix in various brain pathologies compared with the routine brain MR imaging protocol.

In this study, EPIMix T1-FLAIR showed excellent anatomic delineation, including differentiation of gray matter–white matter, comparable with that of the routine sequence. However, geometric distortion, especially in the skull base and frontal lobe, downgraded the overall image quality of EPIMix T1-FLAIR. Although a higher acceleration factor reduces geometric distortion and T2* blurring on images with increasingly effective image sharpness, the susceptibility artifacts of EPIMix could not be completely overcome and there remained some g-factor-related SNR loss.21 Additionally, EPIMix T2-FLAIR showed good overall image quality with good-to-excellent anatomic delineation. Our findings are consistent with those of previous studies that showed image quality and diagnostic performance comparable with that of conventional T2-FLAIR in patients with acute stroke,24 and the contrast-to-noise ratio of EPI-FLAIR was also comparable with that of conventional FLAIR.25 However, EPI T2-FLAIR may be particularly useful in only limited clinical situations in the acute setting because it involves inherently unavoidable susceptibility artifacts. On EPIMix T2WI, the overall image quality and anatomic delineation were far inferior to those of the other EPIMix sequences because T2WI is basically a b=0 image. Therefore, it is essential to improve the image quality of T2WI to broaden the clinical use of EPIMix. Furthermore, EPIMix has a well-known limitation of unavoidable geometric distortion and susceptibility artifacts in all sequences, especially in the skull base, frontal lobe, posterior fossa, and brain stem. These artifacts may limit diagnostic performance in these regions and render EPIMix unsuitable for patients undergoing stereotactic surgery or radiation therapy and for postoperative follow-up imaging.21

Despite these shortcomings, EPIMix can generate all 5 essential sequences in 1 minute with minimal bore time for the patients. Therefore, the shorter scan time allows EPIMix to be used as a motion-resistant MR imaging protocol in certain clinical situations such as for use with patients prone to motion, pediatric or pregnant patients, or patients with claustrophobia or pain, by decreasing patient anxiety and reducing the need for sedation. It also has superior soft-tissue contrast compared with CT without generating radiation during the examination, and EPIMix DWI has excellent image quality, comparable with that of routine DWI. Therefore, EPIMix can be a useful option in patients with time-critical diseases. Furthermore, if EPIMix is run at the beginning of an examination, it may also help in the planning of detailed conventional brain MR imaging sequences.

This study has several limitations. First, it was retrospective and may have, therefore, involved unavoidable selection bias. Second, we did not perform quantitative analysis such as of the contrast-to-noise ratio or SNR because the shorter scan time inevitably leads to degraded overall image quality as also reported by previous studies.15,19,20 Although the image quality analysis is limited by the reader’s subjective judgment, the readers could intuitively perceive the effect of the contrast-to-noise ratio and SNR in the 2 brain MR imaging protocols during the image quality analysis. Third, we could not randomize the acquisition order of EPIMix and routine MR imaging because of the retrospective study design. Fourth, we did not perform a direct comparison between the EPIMix protocol and brain CT in this study. Further comparative studies with head CT would be helpful to confirm the clinical utility of our EPIMix protocol as a proper CT substitute. However, further studies need to be carefully designed because CT involves radiation exposure. Fifth, interobserver agreement between the 2 readers was provided as a percentage agreement because κ statistics provided a paradoxically low value due to an imbalance in the number of concordant and discordant pairs.26,27 Last, EPIMix is a noncommercialized sequence under technical development. It is, therefore, challenging to immediately apply it on different MR imaging scanners from multiple vendors.

CONCLUSIONS

The overall EPIMix image quality was sufficient for diagnostic use with fewer motion artifacts. However, EPIMix is limited by its inherent geometric distortion, being particularly problematic in the skull base near the tissue-air boundaries, and has lower resolution compared with fast spin-echo-based series in a routine protocol. Therefore, it cannot completely replace the routine MR imaging protocol at this time; however, it may be a feasible option in specific clinical situations such as screening, time-critical disease, or patients prone to motion. After further development of the improved EPIMix version, future studies with a larger sample size will help expand the clinical indications of EPIMix and validate our results.

Footnotes

  • Disclosures: Tim Sprenger—UNRELATED: Employment: GE Healthcare, Comments: Tim Sprenger is one of the joint developers of the multicontrast EPI sequences with Professor Skare, and they provided the multicontrast EPI research pulse sequence for this study. However, GE Healthcare had no role in the design of the study, data analyses, or data interpretation.

Indicates open access to non-subscribers at www.ajnr.org

References

  1. 1.↵
    1. Heidemann RM,
    2. Ozsarlak O,
    3. Parizel PM, et al
    . A brief review of parallel magnetic resonance imaging. Eur Radiol 2003;13:2323–37 doi:10.1007/s00330-003-1992-7 pmid:12942278
    CrossRefPubMed
  2. 2.↵
    1. Deshmane A,
    2. Gulani V,
    3. Griswold MA, et al
    . Parallel MR imaging. J Magn Reson Imaging 2012;36:55–72 doi:10.1002/jmri.23639 pmid:22696125
    CrossRefPubMed
  3. 3.↵
    1. Barth M,
    2. Breuer F,
    3. Koopmans PJ, et al
    . Simultaneous multislice (SMS) imaging techniques. Magn Reson Med 2016;75:63–81 doi:10.1002/mrm.25897 pmid:26308571
    CrossRefPubMed
  4. 4.↵
    1. Warntjes JB,
    2. Leinhard OD,
    3. West J, et al
    . Rapid magnetic resonance quantification on the brain: Optimization for clinical usage. Magn Reson Med 2008;60:320–29 doi:10.1002/mrm.21635 pmid:18666127
    CrossRefPubMed
  5. 5.↵
    1. Gulani V,
    2. Schmitt P,
    3. Griswold MA, et al
    . Towards a single-sequence neurologic magnetic resonance imaging examination: multiple-contrast images from an IR TrueFISP experiment. Invest Radiol 2004;39:767–74 doi:10.1097/00004424-200412000-00008 pmid:15550838
    CrossRefPubMed
  6. 6.↵
    1. Hagiwara A,
    2. Warntjes M,
    3. Hori M, et al
    . SyMRI of the brain: rapid quantification of relaxation rates and proton density, with synthetic MRI, automatic brain segmentation, and myelin measurement. Invest Radiol 2017;52:647–57 doi:10.1097/RLI.0000000000000365 pmid:28257339
    CrossRefPubMed
  7. 7.↵
    1. Ashley WW Jr.,
    2. McKinstry RC,
    3. Leonard JR, et al
    . Use of rapid-sequence magnetic resonance imaging for evaluation of hydrocephalus in children. J Neurosurg 2005;103:124–30 doi:10.3171/ped.2005.103.2.0124 pmid:16370277
    CrossRefPubMed
  8. 8.↵
    1. Niederhauser BD,
    2. McDonald RJ,
    3. Eckel LJ, et al
    . Retrospective review of rapid pediatric brain MR imaging at an academic institution including practice trends and factors affecting scan times. AJNR Am J Neuroradiol 2013;34:1836–40 doi:10.3174/ajnr.A3510 pmid:23557956
    Abstract/FREE Full Text
  9. 9.↵
    1. Ryan ME,
    2. Jaju A,
    3. Ciolino JD, et al
    . Rapid MRI evaluation of acute intracranial hemorrhage in pediatric head trauma. Neuroradiology 2016;58:793–99 doi:10.1007/s00234-016-1686-x pmid:27072355
    CrossRefPubMed
  10. 10.↵
    1. Wagner MW,
    2. Kontzialis M,
    3. Seeburg D, et al
    . Acute brain imaging in children: can MRI replace CT as a screening tool? J Neuroimaging 2016;26:68–74 doi:10.1111/jon.12310 pmid:26514114
    CrossRefPubMed
  11. 11.↵
    1. Kralik SF,
    2. Yasrebi M,
    3. Supakul N, et al
    . Diagnostic performance of ultrafast brain MRI for evaluation of abusive head trauma. AJNR Am J Neuroradiol 2017;38:807–13 doi:10.3174/ajnr.A5093 pmid:28183837
    Abstract/FREE Full Text
  12. 12.↵
    1. Missios S,
    2. Quebada PB,
    3. Forero JA, et al
    . Quick-brain magnetic resonance imaging for nonhydrocephalus indications. J Neurosurg Pediatr 2008;2:438–44 doi:10.3171/PED.2008.2.12.438 pmid:19035694
    CrossRefPubMed
  13. 13.↵
    1. Penzkofer AK,
    2. Pfluger T,
    3. Pochmann Y, et al
    . MR imaging of the brain in pediatric patients: diagnostic value of HASTE sequences. AJR Am J Roentgenol 2002;179:509–14 doi:10.2214/ajr.179.2.1790509 pmid:12130464
    CrossRefPubMed
  14. 14.↵
    1. U-King-Im JM,
    2. Trivedi RA,
    3. Graves MJ, et al
    . Utility of an ultrafast magnetic resonance imaging protocol in recent and semi-recent strokes. J Neurol Neurosurg Psychiatry 2005;76:1002–05 doi:10.1136/jnnp.2004.046201 pmid:15965212
    Abstract/FREE Full Text
  15. 15.↵
    1. Nael K,
    2. Khan R,
    3. Choudhary G, et al
    . Six-minute magnetic resonance imaging protocol for evaluation of acute ischemic stroke: pushing the boundaries. Stroke 2014;45:1985–91 doi:10.1161/STROKEAHA.114.005305 pmid:24916906
    Abstract/FREE Full Text
  16. 16.↵
    1. Tanenbaum LN,
    2. Tsiouris AJ,
    3. Johnson AN, et al
    . Synthetic MRI for clinical neuroimaging: results of the Magnetic Resonance Image Compilation (MAGiC) prospective, multicenter, multireader trial. AJNR Am J Neuroradiol 2017;38:1103–10 doi:10.3174/ajnr.A5227 pmid:28450439
    Abstract/FREE Full Text
  17. 17.↵
    1. Ryu KH,
    2. Baek HJ,
    3. Moon JI, et al
    . Initial clinical experience of synthetic MRI as a routine neuroimaging protocol in daily practice: a single-center study. J Neuroradiol 2019 Apr 2. [Epub ahead of print] doi:10.1016/j.neurad.2019.03.002 pmid:30951770
    CrossRefPubMed
  18. 18.↵
    1. Vargas MI,
    2. Boto J,
    3. Delatre BM
    . Synthetic MR imaging sequence in daily clinical practice. AJNR Am J Neuroradiol 2016;37:E68–E69 doi:10.3174/ajnr.A4895 pmid:27444944
    FREE Full Text
  19. 19.↵
    1. Prakkamakul S,
    2. Witzel T,
    3. Huang S, et al
    . Ultrafast brain MRI: clinical deployment and comparison to conventional brain MRI at 3T. J Neuroimaging 2016;26:503–10 doi:10.1111/jon.12365 pmid:27273370
    CrossRefPubMed
  20. 20.↵
    1. Ryu KH,
    2. Choi DS,
    3. Baek HJ, et al
    . Clinical feasibility of 1-min ultrafast brain MRI compared with routine brain MRI using synthetic MRI: a single center pilot study. J Neurol 2019;266:431–39 doi:10.1007/s00415-018-9149-4 pmid:30536110
    CrossRefPubMed
  21. 21.↵
    1. Skare S,
    2. Sprenger T,
    3. Norbeck O, et al
    . A 1-minute full brain MR exam using a multicontrast EPI sequence. Magn Reson Med 2018;79:3045–54 doi:10.1002/mrm.26974 pmid:29090483
    CrossRefPubMed
  22. 22.↵
    1. Delgado AF,
    2. Kits A,
    3. Bystam J, et al
    . Diagnostic performance of a new multicontrast one-minute full brain exam (EPIMix) in neuroradiology: a prospective study. J Magn Reson Imaging 2019;50:1824–33 doi:10.1002/jmri.26742 pmid:30932287
    CrossRefPubMed
  23. 23.↵
    1. Likert R
    . A technique for the measurement of attitudes. Archives of Psychology 1932;22:1–55 https://psycnet.apa.org/record/1933-01885-001. Accessed July 28, 2019
  24. 24.↵
    1. Meshksar A,
    2. Villablanca JP,
    3. Khan R, et al
    . Role of EPI-FLAIR in patients with acute stroke: a comparative analysis with FLAIR. AJNR Am J Neuroradiol 2014;35:878–83 doi:10.3174/ajnr.A3786 pmid:24335543
    Abstract/FREE Full Text
  25. 25.↵
    1. Korogi Y,
    2. Sugahara T,
    3. Shigematsu Y, et al
    . Ultrafast FLAIR imaging with single-shot echo-planar technique in evaluation of intracranial lesions. Comput Med Imaging Graph 1999;23:119–26 doi:10.1016/S0895-6111(99)00008-7 pmid:10397354
    CrossRefPubMed
  26. 26.↵
    1. Viera AJ,
    2. Garrett JM
    . Understanding interobserver agreement: the kappa statistic. Fam Med 2005;37:360–63 pmid:15883903
    PubMed
  27. 27.↵
    1. Feinstein AR,
    2. Cicchetti DV
    . High agreement but low kappa, I: the problems of two paradoxes. J Clin Epidemiol 1990;43:543–49 doi:10.1016/0895-4356(90)90158-L pmid:2348207
    CrossRefPubMed
  • Received October 4, 2019.
  • Accepted after revision January 2, 2020.
  • © 2020 by American Journal of Neuroradiology
View Abstract
PreviousNext
Back to top

In this issue

American Journal of Neuroradiology: 41 (3)
American Journal of Neuroradiology
Vol. 41, Issue 3
1 Mar 2020
  • Table of Contents
  • Index by author
  • Complete Issue (PDF)
Advertisement
Print
Download PDF
Email Article

Thank you for your interest in spreading the word on American Journal of Neuroradiology.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Clinical Experience of 1-Minute Brain MRI Using a Multicontrast EPI Sequence in a Different Scan Environment
(Your Name) has sent you a message from American Journal of Neuroradiology
(Your Name) thought you would like to see the American Journal of Neuroradiology web site.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Cite this article
K.H. Ryu, H.J. Baek, S. Skare, J.I. Moon, B.H. Choi, S.E. Park, J.Y. Ha, T.B. Kim, M.J. Hwang, T. Sprenger
Clinical Experience of 1-Minute Brain MRI Using a Multicontrast EPI Sequence in a Different Scan Environment
American Journal of Neuroradiology Mar 2020, 41 (3) 424-429; DOI: 10.3174/ajnr.A6427

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
0 Responses
Respond to this article
Share
Bookmark this article
Clinical Experience of 1-Minute Brain MRI Using a Multicontrast EPI Sequence in a Different Scan Environment
K.H. Ryu, H.J. Baek, S. Skare, J.I. Moon, B.H. Choi, S.E. Park, J.Y. Ha, T.B. Kim, M.J. Hwang, T. Sprenger
American Journal of Neuroradiology Mar 2020, 41 (3) 424-429; DOI: 10.3174/ajnr.A6427
del.icio.us logo Twitter logo Facebook logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Purchase

Jump to section

  • Article
    • Abstract
    • ABBREVIATIONS:
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • CONCLUSIONS
    • Footnotes
    • References
  • Figures & Data
  • Supplemental
  • Info & Metrics
  • Responses
  • References
  • PDF

Related Articles

  • PubMed
  • Google Scholar

Cited By...

  • Tailored Magnetic Resonance Fingerprinting
  • Crossref (19)
  • Google Scholar

This article has been cited by the following articles in journals that are participating in Crossref Cited-by Linking.

  • Predicting Hypoperfusion Lesion and Target Mismatch in Stroke from Diffusion-weighted MRI Using Deep Learning
    Yannan Yu, Soren Christensen, Jiahong Ouyang, Fabien Scalzo, David S. Liebeskind, Maarten G. Lansberg, Gregory W. Albers, Greg Zaharchuk
    Radiology 2023 307 1
  • One‐Minute Multi‐contrast Echo Planar Brain MRI in Ischemic Stroke: A Retrospective Observational Study of Diagnostic Performance
    Annika Kits, Francesca De Luca, Jens Kolloch, Susanne Müller, Michael V. Mazya, Stefan Skare, Anna Falk Delgado
    Journal of Magnetic Resonance Imaging 2021 54 4
  • A 78 Seconds Complete Brain MRI Examination in Ischemic Stroke: A Prospective Cohort Study
    Siri af Burén, Annika Kits, Lucas Lönn, Francesca De Luca, Tim Sprenger, Stefan Skare, Anna Falk Delgado
    Journal of Magnetic Resonance Imaging 2022 56 3
  • Is improved access to magnetic resonance imaging imperative for optimal ischemic stroke care?
    Mukesh Kumar, Sherry Hu, Steven Beyea, Noreen Kamal
    Journal of the Neurological Sciences 2023 446
  • Optimization of magnetization transfer contrast for EPI FLAIR brain imaging
    Serdest Demir, Bryan Clifford, Wei‐Ching Lo, Azadeh Tabari, Augusto Lio M. Goncalves Filho, Min Lang, Stephen F. Cauley, Kawin Setsompop, Berkin Bilgic, Michael H. Lev, Pamela W. Schaefer, Otto Rapalino, Susie Y. Huang, Tom Hilbert, Thorsten Feiweier, John Conklin
    Magnetic Resonance in Medicine 2022 87 5
  • Rapid processing and quantitative evaluation of structural brain scans for adaptive multimodal imaging
    František Váša, Harriet Hobday, Ryan A. Stanyard, Richard E. Daws, Vincent Giampietro, Owen O'Daly, David J. Lythgoe, Jakob Seidlitz, Stefan Skare, Steven C. R. Williams, Andre F. Marquand, Robert Leech, James H. Cole
    Human Brain Mapping 2022 43 5
  • Ultrafast lumbar spine MRI protocol using deep learning–based reconstruction: diagnostic equivalence to a conventional protocol
    Masahiro Fujiwara, Nobuo Kashiwagi, Chisato Matsuo, Hitoshi Watanabe, Yoshimori Kassai, Atsushi Nakamoto, Noriyuki Tomiyama
    Skeletal Radiology 2023 52 2
  • Tissue volume estimation and age prediction using rapid structural brain scans
    Harriet Hobday, James H. Cole, Ryan A. Stanyard, Richard E. Daws, Vincent Giampietro, Owen O’Daly, Robert Leech, František Váša
    Scientific Reports 2022 12 1
  • Current Perspectives of Artificial Intelligence in Pediatric Neuroradiology: An Overview
    Dann Martin, Elizabeth Tong, Brendan Kelly, Kristen Yeom, Vivek Yedavalli
    Frontiers in Radiology 2021 1
  • Impacts of land use land cover change on Leopard ( Panthera pardus ) habitat suitability and its effects on human wildlife conflict in Hirkiso forest, Sibu Sire District, Western Ethiopia
    Mengistu Muleta Gurmessa, Mitiku Badasa Moisa, Gutema Jira Tesso, Mosissa Geleta Erena
    All Earth 2024 36 1

More in this TOC Section

  • Diagnostic Neuroradiology of Monoclonal Antibodies
  • Cerebral ADC Changes in Fabry Disease
  • ML for Glioma Molecular Subtype Prediction
Show more Adult Brain

Similar Articles

Advertisement

Indexed Content

  • Current Issue
  • Accepted Manuscripts
  • Article Preview
  • Past Issues
  • Editorials
  • Editors Choice
  • Fellow Journal Club
  • Letters to the Editor

Cases

  • Case Collection
  • Archive - Case of the Week
  • Archive - Case of the Month
  • Archive - Classic Case

Special Collections

  • Special Collections

Resources

  • News and Updates
  • Turn around Times
  • Submit a Manuscript
  • Author Policies
  • Manuscript Submission Guidelines
  • Evidence-Based Medicine Level Guide
  • Publishing Checklists
  • Graphical Abstract Preparation
  • Imaging Protocol Submission
  • Submit a Case
  • Become a Reviewer/Academy of Reviewers
  • Get Peer Review Credit from Publons

Multimedia

  • AJNR Podcast
  • AJNR SCANtastic
  • Video Articles

About Us

  • About AJNR
  • Editorial Board
  • Not an AJNR Subscriber? Join Now
  • Alerts
  • Feedback
  • Advertise with us
  • Librarian Resources
  • Permissions
  • Terms and Conditions

American Society of Neuroradiology

  • Not an ASNR Member? Join Now

© 2025 by the American Society of Neuroradiology All rights, including for text and data mining, AI training, and similar technologies, are reserved.
Print ISSN: 0195-6108 Online ISSN: 1936-959X

Powered by HighWire